Menu description
There are 5 items in the main menu: 1 parameter setting, 2 instrument setting and 3 functional testing
4 Data printing 5 Startup test Figure 1
* The menu is a flat touch menu, which can be accessed by direct touch option.
(I) Parameter setting.
The test parameters include 8 options: See Figure 2
Figure 2
1. Program sequence number
The device can prestore 16 sets of control procedures (1--16).
Used to set up different test programs.
Use arrow keys (up and down arrows) to select the program number.
2. Sample name
Name or code for sample input (note: Numbers, English and Chinese are allowed)
3. Initial distillation power
Used to set the heating power before the initial distillation point.
After tapping the input options, the white input box will pop up the keyboard, press the number key, and press the data
Ok key to complete input
4. Final distillation power
Used to set the heating power from the recovery volume to 94% to the final distillation point.
After tapping the input options, the white input box will pop up the keyboard, press the number key, and press the data
Ok key to complete input
5. Condensation temperature
Used to set the constant temperature of the coolant.
After tapping the input options, the white input box will pop up the keyboard, press the number key, and press the data
Ok key to complete input
6, stop volume
Used to set the maximum volume in the recovery cylinder.
After tapping the input options, the white input box will pop up the keyboard. Press the number key for data and press the ok key to complete the input
7. Termination temperature
To prevent overheating from causing danger.
The second term is used to set the maximum temperature of the steam during the experiment.
After tapping the input options, the white input box will pop up the keyboard. Press the number key for data and press the ok key to complete the input
Note: The above parameters must be set according to save, otherwise the data will not be saved