The concept of initial distillation point and dry point and the instrument technical parameters

November 25, 2021
Latest company news about The concept of initial distillation point and dry point and the instrument technical parameters

The concept of initial distillation point and dry point and the instrument technical parameters

 

latest company news about The concept of initial distillation point and dry point and the instrument technical parameters  0

 

The range of boiling points obtained by distilling petroleum under standard conditions is called the "distillation range". It is a label of the amount and temperature of oil that may be distilled from the petroleum product within a certain temperature range.
Distillation range refers to the distillation of oil under specified conditions, from the initial distillation point to the final distillation point of evaporation characteristics of the temperature range.
The whole distillation process includes the items of initial distillation point and final distillation point (or dry point), gasoline also includes 10%, 50%, 90% distillation temperature, residual amount, loss amount of these three items.
Under the specified test conditions, 100ml oil is heated and distilled, and the temperature at which the first drop of oil is extracted from the end of the condenser is called the initial distillation point. In the final stage of distillation, the highest temperature at which all the liquid has evaporated is called the final distillation point. During the test, when the distilled condensate recovered from the measuring cylinder is 10mL, 50ml and 90ml, the distillation temperature is called 10%, 50% and 90% respectively.
Dry point: The highest temperature at which the mercury column of a thermometer stops rising and begins to fall with continued heating, when the distillation range of an oil is determined under specified conditions.
SH6536 automatic distillation tester (Shandong Shengtai Instrument Co., LTD.) adopts color LIQUID crystal display screen, Chinese menu man-machine dialogue, guided operation, the determination process is all automated. It is designed according to the requirements of national standard GB/T6536 "Petroleum Products Distillation Determination method" and ASTMD 86. It adopts embedded computer, imported industrial computer, 12.1 inch LCD touch screen, friendly man-machine dialogue interface and convenient test operation. Using single chip microcomputer and intelligent control technology, the distillation test process of heating and cooling, distillation liquid level tracking, distillation rate control, recovery liquid volume control, temperature recording, test results printing, all automatically completed.


Main performance features
1, real-time display of steam temperature, heater temperature, distillation rate of the whole process of the test three curves, real-time judge whether the relevant indicators meet the requirements of the test standard.
2, the use of imported compressor, imported temperature detection sensor, automatic high subdivision step liquid level tracking system and other key components, accurate and reliable test results, good repeatability.
3, according to user needs and test oil samples, add applicable standards and distillation procedures, and can be called at any time.
4, with atmospheric pressure automatic detection function, automatically complete pressure correction, test results are not affected by atmospheric changes.
5, after the test, the input residual amount can automatically calculate the evaporation temperature.
6, cold bath filling antifreeze, built-in circulation stirring, equipped with liquid level sensor and overflow pipe, keep the water level in the bath normal.
7, equipped with flame detector, with automatic fire extinguishing function, to ensure the safety of the instrument.


Main technical indicators and parameters
1, cold bath temperature control range: (0 ~ 60)℃
2, cold bath temperature control precision: ±0.5℃;
3. Temperature control range of the receiving room: (0 ~ 60)℃;
4, receiving room temperature control precision: ±1℃;
5, distillation heater: 1000W, 24V;
6, distillation liquid product detection range: (0 ~ 100)mL, resolution 0.01mL;
7, distillation liquid product measurement accuracy: ≤ 0.1ml;
8, working power supply: AC(220±10%) V, 50Hz;
9, power consumption: 2500W;
10, the use of environment: ambient temperature (10 ~ 35)℃; Relative humidity ≤80%;
11, dimensions: 500㎜ * 530㎜ * 660㎜ (length × width × height);
12, host weight: 85kg.